When every minute counts

Fastest diagnosis on the spot for patients with acute coronary syndrome

SPECIFICATIONS

Method: rapid determination of cardiac markers troponin T, myoglobin and D-dimer from a single whole blood samples - reflection-photometric method
Measuring time: 8- 12 minutes -
Sample volume: 150 ml -
Working conditions: temperature 18 - 35 ° C; relative humidity 20 - 80% -
Memory: 20 values with time, date, patient data and the user, control 20 reading, 62 IQC results, 15 lot codes -
Dimensions: 173.2 x 124.5 x 44.5 mm -
Weight: 0,454 kg -
Power: 100- 240; 50-60 Hz -

 

: Tests


: Cardiac M Myoglobin
 Quantitative determination of myoglobin in venous whole blood -
 The reaction time 8 minutes -
 Measuring range: 30-700 ng / ml -
 Cut-off: 70 mg / ml -

 
: MYoglobin
is an earlier indicator of acute myocardial infarction -
enables the detection of acute myocardial infarction within 6 hours after the onset of symptomsa -

: Cardiac T Quantitive
Quantitative determination of cardiac troponin from heparinized venous whole blood -
Reaction time: 12 minutes -
Measuring range 0.1-2.0 ng / ml -
Cardiac reader reported results between 0.03 ng / ml and 0.1 ng / ml -
Cut-off: 0.1 ng / ml -

: Troponin T
Is specific to cardiac muscle -
Offers wide possibilities of diagnostics -
Confirmation of suspected myocardial infarction -
Establishment of risk in unstable angina pectoris -
Detection of minor myocardial damage -
Detection of myocardial damage even 10-14 days after emergence -


: Cardiac D-Dimer
 Quantitative determination of D-dimer from heparinized venous whole blood -
 Reaction time: 8 minutes -
 Measuring range: 0.1 to 3.0 mg / ml -
 Cut-off: 0,5μg / ml -

: D- dimer
indicates the occurrence of deep vein thrombosis and pulmonary embolism -

: Cardiac proBNP
Quantitative determination of NT proBNP from heparinized venous whole blood -
Reaction time: 12 minutes -
Measuring range: 60 to 3000 pg / ml -


Easy to use
.The instrument guides you through the measurement process in steps
Insert a test strip into the instrument and press START 1
Apply 150 μl of heparinized whole venous blood on the strip 2
When the reaction is complete, read the results (the results can also be printed using the appropriate printer) 3

 

Cardiac Reader analyzer for quantitative determination of troponin T, myoglobin and D-dimer
Compact, portable instrument that does not take much space -
Always ready for use (stand-by-mode) -
No manual calibration -
All the necessary information concerning the measurement operation are entered automatically using the code chip -
No additional processing of the sample -

PRODUCTS

Product categories that you can find here

Diabetes

Products and equipment of superior quality for all types of diabetes.

Clinical chemistry

All types of analyzers, instruments and products for clinical chemistry.

Medical equipment

List of medical equipment for the practice of family and general medicine.

Nutritional Supplements

Specialized program of natural food supplements.

  • Antuna Hangija 35, 71 000 Sarajevo
  • 387-33-922-244+ ; 387-33-204-696+
  • 387-33-922-594+
  • www.sanam.ba

Sanam Styrka

Send us a message

  Mail is not sent.   Your email has been sent.
Top